Chamber Specifications
Chamber Specifications
Name | Dimensions |
---|---|
Chamber Dimensions | 11.58 m x 7.29 m 7.36 m (interior, with 0.4 m raised floor) |
Wall Shielding | Magnetic field: 20 dB @ 1 kHz, 56 dB @ 10 kHz, 100 dB @ 200 kHz Electric field: 100 dB from 200 kHz to 50 MHz Plane wave: 100 dB from 50 MHz to 1 GHz Microwave: 100 dB @ 18 GHz |
Door Shielding | Magnetic field: 20 dB @ 1 kHz, 70 dB @ 14 kHz, 100 dB @ 200 kHz Electric field: 90 dB @ 1 kHz, 120 dB from 10 kHz to 50 MHz Plane wave: 120 dB from 50 MHz to 10 GHz Microwave: 100 dB from 10 GHz to 26 GHz, 80 dB @ 40 GHz |
Absorber | Power handling: 775 W/m2 Attenuation: >10 dB from 30 MHz to 18 GHz |
Further Specifications
Name | Description |
---|---|
Ferrite Tiles | Handles up to 200 V/m of continuous RF field |
Site-VSWR | Quiet zone test volume: 1.5 m Test frequency range: 1 GHz – 18 GHz Measured performance: 3.37 dB Satisfied CISPR 16-1-4 performance requirement ≤ 6 dB |
Field Uniformity | Test plane: 1.5 m x 1.5 m Test frequency range: 80 MHz to 1 GHz Field level evaluated: 5.4 V/m Measured performance: < 4.2 dB Satisfied IEC 61000-4-3 standard ≤ 6 dB |
Free-space Normalized Site Attenuation | Test range: 3 m Test diameter volume: 1.5 m Measured performance: ± 3.9 dB deviation from theoretical value specified in CISPR 16-1-4 (Ed.3) |
CTIA Ripple Testing | Eligible for CTIA Authorized Testing Laboratory (CATL) filing Test data available upon request |
Connector Pane | 3 N-type connectors 6 FSMA penetration kit 6 FC-type optical connectors 2 K-type connectors 2 ST penetration kit |